Problem Cause Solution
What can I do if... The selected temperature is too high – Reduce the temperature
and the sample oxidizes Put a glass fiber filter on top
of the sample
Sample boils or scorches Reduce the sample quantity
and the splashes continuously or apply the sample more evenly
change the sample weight Select a semi-automatic shutoff
parameter or select the timer mode
Analysis time is too long Increase the temperature
Reduce the sample quantity
Preheat the unit by running it for 2 or
3 minutes with an empty sample pan
Sample loses weight Remove the sample pan and apply
before it is analyzed the sample outside the sample chamber
Sample is liquid or pasty Use a glass fiber filter
Sample contains too little moisture Increase the sample quantity
Insufficient heat output Clean the protective glass or
temperature sensor
Place of installation exposed to Change the place of installation
interfering factors (vibration, etc.)
If any other errors occur, please contact your local Sartorius Service Center