EasyManuals Logo

Sartorius MIS1 Service Manual

Sartorius MIS1
52 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #42 background imageLoading...
Page #42 background image
42
Adjusting the Weighing Platform
Adjustment
resistor
(in ohm)
Test weight
50kg
Test weight
100kg
Test weight
200kg
Test weight
500kg
Test weight
1000kg
0
0.00 - 0.02 0.00 - 0.05 0.0 - 0.1 0.0 - 0.1 0.0 - 0.3
0.39
0.02 - 0.04 0.05 - 0.07 0.1 - 0.1 0.1 - 0.4 0.3 - 0.7
0.82
0.04 - 0.06 0.08 - 0.11 0.1 - 0.2 0.4 - 0.5 0.8 - 1.1
1.2
0.06 - 0.07 0.12 - 0.15 0.2 - 0.3 0.6 - 0.8 1.2 - 1.5
1.6
0.08 - 0.09 0.16 - 0.19 0.3 - 0.4 0.8 - 1.0 1.6 - 1.9
2
0.1 - 0.13 0.20 - 0.26 0.4 - 0.5 1.0 - 1.3 2.0 - 2.6
2.7
0.13 - 0.14 0.27 - 0.29 0.5 - 0.6 1.3 - 1.5 2.7 - 2.9
3
0.15 - 0.16 0.30 - 0.32 0.6 - 0.7 1.5 - 1.6 3.0 - 3.2
3.3
0.16 - 0.19 0.33 - 0.38 0.7 - 0.8 1.6 - 1.9 3.3 - 3.8
3.9
0.19 - 0.23 0.39 - 0.46 0.8 - 0.9 1.9 - 2.3 3.9 - 4.6
4.7
0.23 - 0.27 0.47 - 0.55 0.9 - 1.1 2.3 - 2.7 4.7 - 5.5
5.6
0.20 - 0.33 0.56 - 0.66 1.1 - 1.3 2.8 - 3.1 5.6 - 6.6
Selecting the Adjustment Resistor
Adjusting Off-center Load with 1000 Ohms Initial Resistance
Off-center (in k
g
)
Important: These are calculated (estimated) values

Table of Contents

Other manuals for Sartorius MIS1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius MIS1 and is the answer not in the manual?

Sartorius MIS1 Specifications

General IconGeneral
BrandSartorius
ModelMIS1
CategoryAccessories
LanguageEnglish

Related product manuals