Repairing the Midrics Indicator
Indicator
Important:
An isolating transformer must be installed between the indicator
and mains before performing work that entails opening the Midrics
indicator housing.
To open the Midrics indicator, remove the four nuts as shown in the illustration on
the left.
Note: After completing maintenance or repair work, check the seal between the
front panel and the housing body for damage and replace if necessary. If
the Midrics indicator in question has an IP65 protection rating,
a special test procedure is used to check the IP65 protection after the
housing has been closed.