Error Codes
Display Cause Solution
Inf 07 Function not allowed in scales Contact your local Service Center
verified for use in legal metrology for details on changing settings
Inf 08 The load on the scale is too heavy to zero Check whether “Tare/zero at power on” is set
the readout (1.12)
Inf 09 Taring is not possible when the gross weight Zero the scale
is a minus value
Inf 22 Error in storing reference value, load is too Put a heavier weight on the scale
light
Inf 23 Error in initializing an application Contact your local Service Center
Inf 29 Minimum load not reached Define a lower value for the minimum load
(in the Application settings, menu item 3.6)
Inf 30-31 Indicator is in XBPI mode Change menu setting or perform “Close” function.
Check the interface and cable.
Contact your local Service Center.
Inf 71 Cannot store the current weight value None
(e.g., if control limits are too low or too high)
Inf 72 Cannot store the current weight value None
(e.g., the transaction counter has reached
its limit)
Inf 73 Data not found or unreadable Contact your local Service Center

Table of Contents

Other manuals for Sartorius MIS1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius MIS1 and is the answer not in the manual?

Sartorius MIS1 Specifications

General IconGeneral
BrandSartorius
ModelMIS1
CategoryAccessories
LanguageEnglish

Related product manuals