Operating Manual EVO1X | LAB1X  17
  Installation 
4  Installation
4.1  Unpacking and Equipment Supplied
 t Open the packaging, making sure to remove all parts carefully.
 t After unpacking the device, check it immediately for any external damage.
 t If you detect any damage, proceed as directed in Chapter “7 Care and 
Maintenance,” page 33.
 t Save the box and all parts of the packaging for any future transport. All cables 
should be unplugged when transporting.
The following parts are included in the equipment supplied:
Model Name PMA.Evolution PMA.HD
Large weighing pan: d 233 mm
x –
Small weighing pan: d 180 mm
– x
USB cable x x
Ex-link converter x x
Potential equalization cable x x
Link cable from converter to scale x x
Installation instructions x x
4.2  Selecting a Setup Location
Select the right setup location: 
 − Set up the device on a stable, even surface that is not exposed to vibrations.
 − Maintain free access to the device at all times.
Choose a location that is not subject to the following negative influences:
 − Heat (heater or direct sunlight)
 − Drafts from open windows, AC systems, and doors
 − Extreme vibrations during weighing
 − Heavy “traffic areas” (personnel)
Acclimatization
Condensation from humidity can form on the surfaces of a cold device when it 
is brought into a warm area. You should therefore let a device that has been 
disconnected from its power source acclimatize for approximately 2 hours before 
reconnecting it to the supply voltage.
Seal on Scales Verified for Use in Legal Metrology:
Legislation requires that a seal be affixed to verified scales. On Sartorius devices, this 
seal takes the form of a sticker with the “Sartorius” logo. If the seal is removed, the 
validity of verification will become void and you must have your scale re-verified. 
The verification supplied here is for verified weighing instruments for use in the EEA. 
Please keep it in a safe place.