Operating Manual EVO1X | LAB1X  21
  Installation 
If deviations are evident during startup due to transport damage (e.g., no display, 
no backlighting), disconnect the scale from the power supply and contact Sartorius 
Service.
Connect the scale to the equipotential bonding conductor using an equipotential 
bonding cable with a gage of at least 4 mm².
 t Connect the cable lug of the equipotential bonding cable to the grounding 
terminal of the scale.
 t Connect the equipotential bonding cable to the customer-supplied equipotential 
bonding conductor.
Connect the ex-link converter to the equipotential bonding conductor using another 
equipotential bonding cable with a gage of at least 4 mm².
 t Connect the cable lug of the equipotential bonding cable to the grounding 
terminal of the ex-link converter.
 t Connect the equipotential bonding cable to the customer-supplied equipotential 
bonding conductor.
Connecting a PC/Notebook
 t Insert a USB cable (1) into the right-hand socket on the ex-link converter and 
connect the cable to a PC or notebook. If the ex-link converter YCO-Y is being 
used and installed in Zone 2 or Division 2 potentially explosive areas and the 
USB cable (1) does not have an intrinsically safe electrical circuit, the cable must 
be secured against disconnection. Refer to the Control Drawing and Safety 
Instructions (Chapter 15, page 42) as well as the following illustrations.
 t Fasten the strain relief YSR01 to the converter (1) using the two screws (2).
1
1
2
2