Profibus Interface PR 5210 Operating Manual
64 Sartorius
6.3.3.12 Parameter P44: Standstill time
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0
Byte 1
Byte 2
Byte 3 STIME
Time expressed in multiples of measuring time at which a standstill is detected.
STIME 1... 9 times measuring time, default = 1
6.3.3.13 Parameter P45: Standstill range
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0
Byte 1
Byte 2 SRANGE MSB
Byte 3 SRANGE LSB
Limit to define the weight to be in standstill.
SRANGE 0 ... 10.00d default = 1d
value = d*100
6.3.3.14 Parameter P46: Standstill timeout
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0
Byte 1
Byte 2
Byte 3 TIME
Number of measuring times during the standstill condition should be reached.
TIME 0 ... 100 M (measuring times), default = 8
If the standstill timeout is exceeded by a function, which needed the standstill condition, the function
is aborted (Taring, set zero, calibrate, 'P'-command of the SMA-protocol). Additional the error code 31
and the 'CmdError' bit is set.
6.3.3.15 Parameter P47: Zeroset range
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0
Byte 1
Byte 2 RANGE MSB
Byte 3 RANGE LSB
RANGE 0 ... 500.00 d default = 50 d
Value = d*100