PR 5210 Operating Manual Profibus Interface
Sartorius 63
6.3.3.8 Parameter P40: Digital filter
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0
Byte 1
Byte 2
Byte 3 Filter
Filter 0 = off, 1 = Bessel, 2 = Aperiodic, 3 = Butterworth, default = off
6.3.3.9 Parameter P41: Filter frequency
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0
Byte 1
Byte 2
Byte 3 FREQ
FREQ Value has to be entered in frequency*100. Example: for 1.56 Hz enter 156. Frequency range
depends on the measuring time, see table in chapter 4.7.2.3
Only valid if the digital filter is not set to on
6.3.3.10 Parameter P42: Measuring time
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0
Byte 1
Byte 2 TIME MSB
Byte 3 TIME LSB
TIME measuring time in in ms 10 ... 1920, default = 320 ms
Only valid if the digital filter is switched to off.
Parameter P53 will be set accordingly.
10...160ms -> P53: 4...0
320...1920ms -> P53: 0
6.3.3.11 Parameter P43: Test mode
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0
Byte 1
Byte 2
Byte 3 MODE
MODE 0 = absolute, 1 = relative, default = 0