15
N
Note: The listed systematic and random error values can be achieved only under strictly controlled conditions during type test per ISO 8655.
Due to the continuous product development by Sartorius, the systematic and random error values are subject to change without prior notice.
Proline
®
Plus FIXED Volume, Single-Channel
Order Code Channels Volume Range Test Volume Systematic Error
N
Limit ±
Random Error
N
Limit
(µl) (µl) (%) (µl) (%) (µl)
728515 1
n
5 5 1.3 0.065 1.2 0.060
728520 1
n
10 10 0.8 0.080 0.8 0.080
728530 1
n
20 20 0.6 0.120 0.5 0.100
728535 1
n
25 25 0.5 0.125 0.3 0.075
728545 1
n
50 50 0.5 0.250 0.3 0.150
728550 1
n
100 100 0.5 0.50 0.3 0.30
728560 1
n
200 200 0.4 0.80 0.2 0.40
728565 1
n
250 250 0.4 1.00 0.2 0.50
728567 1
n
500 500 0.3 1.50 0.2 1.00
728570 1
n
1,000 1,000 0.3 3.0 0.2 2.0
728575 1
n
2,000 2,000 0.3 6.0 0.15 3.0
728580 1
n
5,000 5,000 0.3 15 0.15 7.5
728590 1
n
10,000 10,000 0.6 60 0.2 20