8
9. Recommendations for Good 
Pipetting
  Make sure that the tip is firmly attached to the tip cone.
  Hold the pipette vertically when aspirating the liquid and place the tip 
only a few millimeters into the liquid.
  Always control the operating button slowly and smoothly.
  Pre-rinse the tip before aspirating the liquid by filling and emptying the 
tip for three to five times. This is important especially when  pipetting 
liquids with a viscosity and density greater than water or liquids with 
high vapor pressure (e.g. ethanol).
  Check that the pipette, tip and liquid are at the same temperature.
  When pipetting liquids with temperatures different to the ambient 
temperature change the tip after each pipetting. Do not pre-rinse the tip. 
  To avoid contamination, do not rest the pipette on its side especially 
with the tip attached.
  Change the tip cone filter regularly (recommendation after 50 - 250 
pipetting cycles).
  Never strike the tip cone against the tip tray when mounting the tip as 
this can damage the pipette.
  Avoid rough handling and do not drop the pipette.
  Avoid exposing the unit to extreme temperature changes, humidity 
and dust (operating temperature from 15°C to 40°C).
10. Maintenance
Proline
® 
Plus pipettes have been designed for easy in-house service. 
If the pipette is used daily it is recommended to clean|decontaminate 
and check the performance of the pipette every three months. Sartorius 
also provides complete repair and recalibration service for your pipette 
including a service report and performance certificate.
Note: Please make sure that the pipette has been decontaminated 
before you send it to us or your local representative. Please advise if any 
hazardous material has been used with your pipette.
Note: The use of the tip cone filters may prolong the service interval. 
Change the filters regularly.
Note: It is recommended to always use gloves when cleaning the 
pipette.
10.1. Daily Cleaning of The Outer Surface of 
The Pipette
Your Proline
® 
Plus pipette should be checked every day for cleanliness. 
To clean and decontaminate the outer surface of your pipette use 
ethanol (70%), isopropanol (60%) or a mild detergent and a soft lint-free 
cloth.
Gently clean the surface of the pipette with moistened cloth and wipe 
dry. Pay special attention to the tip cone. Change the tip cone filter if 
needed (Chapter 7.3.)