Specifications
User Manual Secura
®
, Quintix
®
, Practum
®
15. Specifications
15.1 General Data
AC adapter
Sartorius AC adapter module 6971790 with replaceable, country-specific power plug adapters
Primary 100 - 240 V~, -10% / +10%, 50 - 60 Hz, 0.2 A
Secondary 15 V DC, ± 5%, 530 mA (max.) / 8 W (max.): 0 to +40°C and 15 V DC, ± 5%,
330 mA (max.) / 5 W (max.): 0 to +50°C
Other data Protection class II in accordance with EN/IEC 60950-1 / up to 3000 m above sea level /
IP40 in accordance with EN 60529/IEC 60529
Balance
Power supply Only via Sartorius AC adapter module 6971790
Input supply voltage 12.0 - 15.0 V DC
Power consumption 2 W (typically)
Ambient Conditions
The specifications apply under the following ambient conditions:
Environment For indoor use only
Ambient temperature* +10°C to +30°C
Operational capability Guaranteed between +5 °C and +45°C
Storage and shipping -10°C to +60°C
Height Up to 3000 m above sea level
Relative humidity** 15% to 80% for temperatures up to 31°C non-condensing,
decreasing linearly up to 50 % relative humidity at 40°C and 20 % at 50°C
Safety of Electrical Equipment In accordance with EN 61010-1/IEC 61010-1 Safety requirements for electrical equipment for
measurement, control and laboratory use – Part 1: General requirements
Electromagnetic Compatibility In accordance with EN 61326-1/IEC 61326-1
Electrical equipment for measurement, control and laboratory use – EMC requirements –
Part 1: General requirements
Interference resistance: Suitable for use in industrial areas
Transient emissions: Class B (suitable for use in residential areas and areas that are connected to a low voltage
network that (also) supplies residential buildings). The device can therefore be used in both
areas.
Standard Equipment
Selectable weight units
1
Gram, kilogram, carat, pound, ounce, troy ounce, Hong Kong tael, Singapore tael, Taiwanese
tael, grain, pennyweight, milligram, parts per pound, Chinese tael, momme, Austrian carat, tola,
baht, mesghal and Newton
Available application programs Secura
®
models: Counting, percentage, density determination, peak hold, unstable condition,
checkweighing, mixing, components (totalizing), statistics, conversion, SQmin, and identifiers
Quintix
®
models: Counting, percentage, density determination, peak hold, unstable condition,
checkweighing, mixing, components (totalizing), statistics, conversion
Practum
®
models: Counting, percentage, density determination, peak hold, unstable condition,
checkweighing
Verified balances in accordance with EU requirements comply with the requirements of Council Directive 2009/23/EC with
EN 45501:1992 and OIML R76:2006.
* For verified balances in accordance with EU requirements, refer to the information on the balance.
** For verified balances in accordance with EU requirements, the legal regulations apply.
1) Depending on the country-specific model version, not all weight units listed may be available.