107
Specifications
User Manual Secura
®
, Quintix
®
, Practum
®
15.3  Verified Models with Country-specific Type Approval Certificate
Secura
®
Model 224-1x 
2
) 124-1x 
2
) 1103-1x 
2
) 613-1x 
2
) 513-1x 
2
) 313-1x 
2
) 213-1x 
2
) 6102-1x 
2
) 5102-1x 
2
) 3102-1x 
2
)
Accuracy class
kkkKKKKKKK
Type 
3
) SQP-A SQP-A SQP-I SQP-B SQP-B SQP-B SQP-B SQP-C SQP-C SQP-C
Max g 220 120 1,100 610 510 310 210 6100 5100 3100
Scale interval d mg 0.1 0.1 11111101010
Verification scale interval e mg 1 1 10 10 10 10 10 100 100 100
Min mg 10 10 100 20 20 20 20 500 500 500
Min (only for Models ...-1OIN)  mg 100 100 – 200 200 200 200 5,000 5,000 5,000
Tare equalization range (subtractive) < 100% of max. weighing capacity
Min. initial weight according to 
USP, typical
g 0.12 0.12 1.5 1.5 1.5 1.5 1.5 12 12 12
Typical stabilization time s 2 2 1.5 1111111
isoCAL: 
– Temperature change 
– Time interval
 
K 
h
 
1.5 
4
 
1.5 
4
 
1.5 
4
 
2 
6
 
2 
6
 
2 
6
 
2 
6
 
2 
6
 
2 
6
 
2 
6
Display result  
(depending on the set filter level)
s 0.2 0.2 0.1/0.2 0.1/0.2 0.1/0.2 0.1/0.2 0.1/0.2 0.1/0.2 0.1/0.2 0.1/0.2
Weighing pan size mm
d 90 d 90
d 120
d 120
d 120 d 120 d 120 d 180 d 180 d 180
Weighing chamber height* mm 209 209 209 209 209 209 209 – – –
Net weight, approx. kg 5.1 5.1
5.9
5.1 5.1 5.1 5.1 5.2 5.2 5.2
Model 2102-1x 
2
) 1102-1x 
2
) 612-1x 
2
) 6101-1x 
2
) 3101-1x 
2
)
Accuracy class
KKKKK
Type 
3
) SQP-C SQP-C SQP-C SQP-C SQP-C
Max g 2,100 1,100 610 6,100 3,100
Scale interval d mg 10 10 10 100 100
Verification scale interval e mg 100 100 100 100 100
Min mg 500 500 500 5,000 5,000
Min (only for Models ...-1OIN) g 55555
Tare equalization range (subtractive) < 100% of max. weighing capacity
Min. initial weight according  
to USP, typical
g1212126060
Typical stabilization time s 11111
isoCAL: 
– Temperature change 
– Time interval
 
K 
h
 
2 
6
 
2 
6
 
2 
6
 
2 
6
 
2 
6
Display result  
(depending on the set filter level)
s 0.1/0.2 0.1/0.2 0.1/0.2 0.1/0.2 0.1/0.2
Weighing pan size mm
d 180 d 180 d 180 d 180 d 180
Net weight, approx. kg 5.2 5.2 5.2 5.2 5.2
2
)  Possible terms for country-specific models:
  x = CEU:   Verified balances with EC Type Approval Certificate D12-09-014 
without country specific additions
  x = CFR:   Verified balances with EC Type Approval Certificate D12-09-014 
for France only
  x = CIT:   Verified balances with EC Type Approval Certificate D12-09-014 
for Italy only
  x = CCH:   Verified balances with EC Type Approval Certificate D12-09-014 
for Switzerland only
  x = CN:  CMC Type Approval Certificate for China
  x = OJP:  Balance with Type Approval Certificate for Japan
  x = OBR:  Balance with Type Approval Certificate for Brazil
  x = ORU: Balance with Type Approval Certificate for Russia
  x = OIN:  Balance with Type Approval Certificate for India
  x = OAU: Balance with Type Approval Certificate for Australia
3
)  All models with “...CN”: type “SQP”
*  Upper edge of the weighing pan to the lower edge of the upper draft shield 
panel