7.2  SQmin Minimum Weighing (Secura
®
 only)
  Purpose:  This function is used to compare the weight value with the defined minimum sample 
quantity (SQmin = sample quantity minimum). This ensures that weighing results are 
above the specified minimum weight defined by your quality assurance system. This 
function is used to observe the minimum weight in order to meet the criteria of the 
United States Pharmacopeia (USP), for example. 
 
SQmin is not the same as the minimum capacity Min in legal metrology. 
 
  Prerequisites:  The balance must be set up by a service technician to be able to use the SQmin 
function. The technician will determine the permitted minimum sample quantity and 
load this to your balance using the guidelines of your QA system. He will document 
this setting via a “Weighing module test as per USP” certificate in which the 
measurements and min. sample quantity are logged. The SQmin function ensures that 
the weighing results correspond to USP guidelines.
Switching SQmin On/Off
If the SQmin function is set up on the balance, go to the 
Settings menu, Weighing – 
SQmin to switch it on/off.
Display of SQmin Function during Weighing
t  Place the sample on the weighing pan.
 
Max 220 g
SQmin 0.1200 g
d
=
0.0001 g
0.1037
+
  y  The SQmin value is displayed in red if the weight value is smaller than the 
specified minimum weight.
The weight value is displayed in gray if it is smaller than the specified SQmin 
value. The weight value cannot be saved or printed out in applications 
(depending on the setting in the “Safety level” menu).