Using the Balance
User Manual Secura
®
, Quintix
®
, Practum
®
5.2.2  Leveling with a Conventional Level Indicator  
(Quintix
®
 and Practum
®
)
L
R
R
L
L
R
  The Quintix
®
 and Practum
®
 models are fitted with a conventional level indicator.  
This is located to the left of the display. The position of the air bubble in the indicator 
shows whether or not the balance is correctly leveled.
t  Twist both of the front leveling feet of the balance to move the air bubble so that 
it is centered in the circular marking.
The figure shows which leveling foot should be twisted in which direction. Generally, 
both feet must be adjusted in order to position the air bubble exactly in the center of 
the circular marking.
Examples:
–  If the air bubble is too far to the lower right, twist the right leveling foot 
counterclockwise (see figure).
–  If the air bubble is too far to the upper right, twist the left leveling foot 
clockwise.
–  If the air bubble is too far to the left, twist the left leveling foot counterclockwise 
and the right leveling foot clockwise.
t  Check the level indicator after each adjustment and correct if necessary.
t  Complete the leveling by calibrating the balance.