System Settings (Setup Menu)
User Manual Secura
®
, Quintix
®
, Practum
®
6.2.7 Identifier (Secura
®
 only)
 
Identifier
Set device ID
Set additional ID
Batch ID function
Sample ID function
  Define identifiers for the printout via the g button and in the ISO/GLP printout here. 
Under Printout in Manual print, the ISO/GLP printout option must be activated. IDs 
can be a maximum 14 characters long.
Set device ID menu option:
The device identification (ID) is printed in the header of the GLP printout.
t  To activate the device ID, select t, enter the desired device ID, and confirm 
withl.
Set Additional ID menu option:
The additional ID (A ID) is also printed in the header of the GLP printout.
t  To activate the additional ID, select t, enter the desired device ID and confirm 
with l.
Batch ID function menu option:
The batch ID (LID) is queried one time at the start of the GLP printout.
t  Activate this option to enter or print out the batch ID.
Sample ID function menu option:
The sample ID (SID) can be activated for each printout by selecting the g button on 
the balance display.
t  Define settings for printing the sample ID.
–  On: The sample ID is activated. This ID is queried before each printout.
–  Autom. increment: Select this option to automatically assign the sample ID 
in ascending order.
–  Autom. decrement: Select this option to automatically assign the sample ID 
in descending order.
–  Off: The sample ID is deactivated.
 
To enter and delete figures and text, e.g., for the sample ID, see “Enter Figures on the 
Keypad” on page 25.