31
System Settings (Setup Menu)
User Manual Secura
®
, Quintix
®
, Practum
®
6.2.3 Device Information
 
Device information
Manufacturer: Sartorius
Model: SECURA2102
Serial number:  0012345678
Version BAC:  00-50-02.xx CN:1701
Version APC:  01-70-02.xx CN:6587
  The manufacturer, model, serial number, and software version of your balance are 
displayed here.
6.2.4 Calibration/Adjustment
 
Calibration/Adjustment
isoCAL
Calibration report
 With Secura
®
 models, it is possible to set how the automatic calibration/adjustment 
function isoCAL is started.
 With Secura
®
 and Quintix
®
 models, saved reports about the most recent calibration 
procedures, including deviations identified, can be viewed and output to a PC or 
Sartorius lab printer.
isoCAL menu option (Secura
®
 only):
t  Select the desired start option for isoCAL.
–  Off: isoCAL is switched off.
–  Info, manual start: A note appears on the display when predefined time 
intervals or temperature values are exceeded. isoCAL can be started 
manually.
–  Automatic: isoCAL is switched on. Calibration/adjustment starts 
automatically when predefined time intervals or temperature values are 
exceeded.
  Calibration report menu option (Secura
®
 and Quintix
®
 only):
t  Select the desired report by date.
If there is more than one calibration report per date, use the < and > buttons to 
navigate through the reports.
t Select a to print the displayed report.
A maximum of 99 entries are saved per day. 
Calibration reports are deleted after 30 days. 
6.2.5 Weighing
 
Weighing
SQmin
Application
Ambient conditions
Safety level
  A number of basic settings for the weighing functions can be defined here. Options 
are limited for verified balances. All settings given in the menu are permitted.
SQmin menu option (Secura
®
 only):
If the SQmin function is set up on the balance, the function can be switched On or 
Off here (see “SQmin Minimum Weighing (Secura
®
 only)” on page 44).
Calibration report
2014-10-04: 2 Protocols
2014-10-01: 2 Protocols
2014-09-25: 3 Protocols
2014-09-24: 3 Protocols