32
System Settings (Setup Menu)
User Manual Secura
®
, Quintix
®
, Practum
®
Safety Level
The Secura
®
 models monitor the following at all times:
–  whether the balance is leveled correctly
– whether adjustment is necessary
–  whether minimum sample quantity requirements are met according to USP.
Safety level menu option (Secura
®
 only):
t  Select the desired safety level for SQmin, isoCAL, and LEVEL
  If a verified balance has not been leveled, the following functions will be disabled 
immediately (as is the case with the “High” safety level): 
– Printout  
–  Startup and saving of application  
Here, it does not matter what settings this menu item has.
 High (factory setting): If one of the three conditions has not been met, a warning is 
displayed straight away. No data is transmitted. 
  The following functions are disabled immediately:
– Printout
–  Startup and saving in applications
Displays:
–  Information about red marking of display element (LEVEL, isoCAL or SQmin)
–  Weight value displayed in gray with a black warning symbol
A warning is displayed after 60 seconds in the event of non-observance.
Standard: If one of the three conditions has not been met, a warning is displayed 
straight away. 
Displays:
–  Information about red marking of display element (LEVEL, isoCAL or SQmin)
–  Weight value displayed in gray with a black warning symbol
A warning is displayed after 60 seconds in the event of non-observance.
Printout:
–  Weight values are marked with “!”
Low: The following information is displayed:
–  SQmin value has not been achieved -> SQmin value is marked red
–  Balance not leveled -> Red marking on “LEVEL” display element
–  Balance not adjusted due to temperature or time -> Red marking on “isoCAL” 
display element
Weighing
SQmin
Application
Ambient conditions
Safety level