USB Port
User Manual Secura
®
, Quintix
®
, Practum
®
Special Outputs
Position 12345678910111213141516
******––******CRLF
or ******Low*****CRLF
or ******High****CRLF
* Space Cal.Ext. External adjustment
– – Final readout Cal.Int. Internal adjustment
Low Underweight CR Carriage return
High Overload LF Line feed
Error message
Position 12345678910111213141516
***Err**/###****CRLF
* Space ### Error number
Example: Output of the weight value + 111.255 g
Position 12345678910111213141516
* * * 111. 255* g* * CRLF
+* * 123. 5[ 6]
1)
g**CRLF
Position 1: Plus +, minus –, or space
Position 2 – 10: Space or weight value with decimal point;
leading zeros are output as spaces.
Position 11: Space
Position 12 – 14: Characters for unit of measure or space
Position 15: Carriage return
Position 16: Line feed
1)
Settings “PC – SBI” and “USB RS232 SBI” for the identification of digits not verified
for use in legal metrology:
In the setting “SBI”, there is no automatic identification of display digits not
verified for use in legal metrology. Corresponding measures or settings must be
carried out on the peripheral device.