16 FH-BA-e-2020
Adjustment instruction FH
1. switch on the
device
Press ON/OFF button
The green light goes on
2. switch to
calibration mode
Directly after pressing the
ON/OFF button, press the
PEAK and PRINT buttons
simultaneously, several times
and very quickly one after the
other until the left red light
comes on.
The left red light goes on
3. device type
Press SET immediately after
the red light has come on.
The maximum Newton
value of the device is
displayed, or can now be
set.
3a) (Back in normal
mode???)
(In the meantime, if you are
back in normal mode, turn off
the power and start over from
step 1. If necessary, press keys
faster)
4. select device
Use the keys ▼▲ to select the
maximum load (N) of the
respective device.
The value matching the
device is shown in the
display.
5. save settings
Press SET
6. switch to
calibration
Press MEMORY
The right red light comes
on
7. indicate the
available calibration
weight
Press UNIT and enter the
calibration weight in Newton
with ▼▲. (X kg * 9.81)
The weight in Newton is
shown in the display
8. save
Press SET and UNIT
simultaneously
9. attach weight
Hang the weight on the device
and keep it as still as possible.
Then press ZERO
The instrument returns to
normal mode and is
calibrated