SAVI Dual™ Migraine Therapy Instructions for Use 15
Delivering the Treatment
NOTE: Two hands are needed to stabilize the device, but only one treatment button
needs to be activated to deliver treatment.
NOTE: The device shuts itself off if treatment is not delivered within 45 seconds. Press
the power button to turn the device back on and restart the treatment procedure.
Once the device is in place and your thumbs are on the treatment switches, press one or both switches and
hold for at least 2 seconds.
The device emits a soft, audible click as the treatment is delivered.
After you have successfully delivered the first pulse, remove the device from your head.
The LED light on the POWER button softly pulses to show that treatment is completed. The device is now
ready for another pulse. You may select to treat again by pressing the power button.
Repeat the subsequent procedures from page 15, Positioning the Device.
After delivering the final pulse and 30 seconds of inactivity, the device will attempt to connect to a cellular
network, transfer treatment history to your SAVI Smart Diary™ and will then power off.
While the power button is lit BLUE place the device on a flat surface at least 8 inches (20 cm) away from you,
other people and electronic equipment.
Delivery of several pulses in a row may cause the device’s surface temperature to exceed 105.8° F (41° C). The
surface may get as warm as 118.4° F (48° C). Let the device cool for 15 minutes. Try again.
Renewing Your SAVI Dual Prescription by Cellular Network
NOTE: The System is intended for use under the care of a doctor. For more
information on renewing your prescription, please contact eNeura Customer Care.
An active prescription is required.
NOTE: The System only delivers treatment if the programmed prescription has not
expired.
The device will display an ORANGE warning light when your prescription is about to expire: Renew your
prescription promptly.
Contact Customer Care to request a refill if you are not enrolled in the SAVI Dual monthly subscription
program.
Upon receipt of payment, eNeura Customer Care will initiate reactivation of your device via the cellular
network.
Ensure the System lock switch is in the OFF position and the AC adapter is NOT connected to your device.
Press the Power Button on your device. After approximately 30 seconds, pulsing blue illumination on the
Power Button indicates the device is connecting to the cellular network to transfer treatment history to
your SAVI Smart Diary or download a new prescription.
After the device has successfully connected to a network, the new Rx data will transfer, and the ORANGE Rx
warning light will become white again. You may resume use.
Contact eNeura Customer Care if you are in a low or no-cell service area or if your device does not connect to
the network. A micro-SIM card can be provided to renew your prescription.