The Free Adjust operating function is included in NSG 438 products from
serial number 521 (March 2006) upwards.
6.3.6. Settings
This branches into a range of sub-menus, thus:
6.3.7. Language
Touch the button and choose the language you wish to use with the wheel.
Note: Language looks like this in Japanese:
6.3.8. Device info
Gives information separately for the pistol and the base station regarding
the version of the equipment and its software.
6.3.9. Discharge
Use the wheel to select either Air-Discharge:
or Contact-Discharge:
The RC values for the relevant network are also shown.
The corresponding value is automatically loaded upon switching between
Air/Contact-Discharge when standard voltage values are selected.
The appropriate type of test finger must, of course, be mounted to suit the
type of discharge:
Air-Discharge = ball-shaped test finger
Contact-Discharge = sharp point test finger
An appropriate error message is displayed if the pistol detects wrong type
of finger has been mounted.