EasyManua.ls Logo

SCHOTT LLS 3 - Page 35

SCHOTT LLS 3
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
User Manual - LLS 3
Page 35
xxxx = command parameter (factory default) = 0000
return value: &pdxxxx where “xxxx” is the Triggered Strobe Delay in microseconds.
xxxii. Get/Set Triggered Strobe On-Time: &poxxxx
xxxx = command parameter
0050 to 9999 is the value in microseconds (decimal)
(factory default) = 0050
return value: &poxxxx where “xxxx” is the Triggered Strobe Delay in microseconds.
xxxiii. Get/Set Triggered Strobe Mode On/Off: &pmx
Activates/Deactivates Triggered Strobe Mode.
x = command parameter
0 = triggered strobe mode off (factory default)
1 = triggered strobe mode on
? = query value
return value: &pmx where “x” is either 0 or 1 depending on whether the light source is in
Triggered Strobe Mode.
xxxiv. Get Device Name Query &q
Obtain device name. Useful for mixed light source technologies residing on the same
network.
return value: &qLED Light Source
xxxv. Get/Set Restore Settings on Power Up: &rx
By default, when the LLS 3 is turned on the settings for the light source is automatically
restored for the LLS 3. This behavior can be disabled with this setting. Use the
Restore Settings command (&t) to restore stored values when &r is set to off
x = command parameter
0 = off (factory default)
1 = on
? = query status
return value: &rx where “x” is either 0 or 1 depending on the setting.
xxxvi. Get/Set Save Settings After Each change: &sx
Excessive writes to the FLASH memory for frequent changes for the light source should
be avoided. The LLS 3 will store settings after 10 seconds of inactive command
changes. The writing to non-volatile memory can be disabled with this setting. Note that
until the Save setting is set back on, the light source will not store changes to its
parameters.
x = command parameter
0 = off (factory default)
1 = on
? = query status
return value: &sx where “x” is either 0 or 1 depending on the setting.
xxxvii. Restore Settings now &t
Restores settings immediately to the LLS 3. This command is useful in conjunction with
the &r command. For example, a user may wish not to restore settings on power
applied but at a specific moment when dictated by the user.
return value: &t
xxxviii. Get Light Source Serial Number &z
Obtains the serial number of the LLS 3 (as indicated on the bottom of the unit).
return value: &zxxxxxx where “xxxxxx” represents a six digit serial number of the LED
light source.