5.5 Print the accumulated output format
NO. 004 NO.
Date: XX.XX.XX (year. month. date)
Time: XX.XX.XX (hour. minute. Second)
Total: 003 (accumulate timesexample for 3 times is 003)
Total.W: 2.88kg accumulate weight
6. Maintenance
6.1 Regular Error and maintain method
Error
Reason instruction
Solution
Display
UUUUUU
1. the loaded weight
excess overload
range of max. capacity
2. wrong connection
with load cell or no
connection with it.
3. load cell
unworkable
1.decrease loaded weight
2. check load cell
connection
3. checking load cell
check input and output
resistance to judge it is
good or not.
Display nnnnnn
1. calibration is no
good
2 cell single line is
connect a wrong line.
3the cell is bad.
1. check scale is resisted
or not, foot is kept level
or not.
2. check load cell
connection.
3. checking load cell
check input and output
resistance to judge it is
good or not.
ERR1
during calibration, no
Input the correct weight