EasyManua.ls Logo

SDS IDS701 - Maintenance; Regular Error and maintain method; Daily maintenance

Default Icon
26 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
IDS701 User Manual
Page 22 of 25
input added weight or
input weight exceed
max capacity.
ERR2
during calibration, the
added weights not
enough
Added weight at least
10%of Max. capacity,
Recommend the weights
is 60-80% the Max.
capacity
ERR3
during calibration,
input single is
negative.
1..Check connection is
correct or not.
1. Check load cell is
damaged or not.
3. renew calibration, if still
wrong. pls replace the
PCB
ERR4
During calibration,
single is unstable
Ensure added weight and
scale is stable, start
calibration
ERR5
EEPROM check error
change PCB.
6.2 Daily maintenance
1. In order to ensure indicator display clearly and prolong use life,
the indicator should not be placed directly on sunlight.
2. Load cell and indicator should be well connected , the system
should have a good ground, away from strong electric field, magnetic
field.
3. Do not use indicator outside in rainy, better keep it power off.
4. Power off firstly while plug and unplug