AG2000 INSTRUCTIONSTROUBLESHOOTING
Problem
Probable Cause Try...
Blank display
Flow rate steadily reads
zero when there is ow
Display reads [ - ]
Flow rate intermittently
drops when there is ow
Jumpy reading
Replace battery pack
Reading will resume when ow increases
Reposition meter for full pipe
Note ow direction arrow, reverse meter
Reposition meter for full pipe or rotate to avoid
bubbles
Check for proper equalization
Use external power source
(allows more ow averaging)
Install chemigation line downstream of meter
(or enough upstream for thorough mixing of
uids before meter)
Dead battery
Flow is below cutoff (very low)
There is air in the meter
Meter is installed backward
There is air in the meter
Improperly equalized
Pulsing ow
Rapidly changing conductivity
(chemigation applications)