Seametrics AG2000 - Troubleshooting; Common Problems and Solutions

Seametrics AG2000
8 pages
Print Icon
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
LT-65200290r1.2 20160729
7/29/16
Seametrics • 19026 72nd Avenue South • Kent, Washington 98032 • USA
(P) 253.872.0284 • (F) 253.872.0285 • 1.800.975.8153 • seametrics.com
AG2000 INSTRUCTIONSTROUBLESHOOTING
Problem
Probable Cause Try...
Blank display
Flow rate steadily reads
zero when there is ow
Display reads [ - ]
Flow rate intermittently
drops when there is ow
Jumpy reading
Replace battery pack
Reading will resume when ow increases
Reposition meter for full pipe
Note ow direction arrow, reverse meter
Reposition meter for full pipe or rotate to avoid
bubbles
Check for proper equalization
Use external power source
(allows more ow averaging)
Install chemigation line downstream of meter
(or enough upstream for thorough mixing of
uids before meter)
Dead battery
Flow is below cutoff (very low)
There is air in the meter
Meter is installed backward
There is air in the meter
Improperly equalized
Pulsing ow
Rapidly changing conductivity
(chemigation applications)