Weighing • 29
2. Attach the transport lock to the lower part of the housing.
4.4 Setting up device
CAUTION!
Faulty measurement as a result of devices being set up incorrectly
Soft floors, such as wooden boards, give under the patient's weight
and falsify the measuring result. Incorrect ambient conditions or lack of
alignment lead to measurement errors.
Select a setup location with a level, stable floor in order to achieve
precise measuring results.
Use the device only in the intended ambient conditions outlined in
the section entitled “Technical data”.
Ensure the device is aligned before using it and every time it
changes location.
Place the device on a firm, level surface.
5 WEIGHING
CAUTION!
Personal injury/faulty measurement
Before using the device each time, perform a function check as de-
scribed in the corresponding section of these instructions for use.
5.1 Adjusting display
To obtain meaningful measured values, you must check before every use that
the pointer of the scale reads “0”.
1. Ensure that there is no load on the scale.
2. Check the current position of the pointer.
3. Move the adjusting wheel so that the pointer reads “0”.
Turn clockwise: Pointer moves counter-clockwise
Turn counter-clockwise: Pointer moves clockwise
17-10-06-389-100h_2024-08S