Troubleshooting • 33
8 TROUBLESHOOTING
If faults occur when operating the device, first attempt to remedy them yourself
using the following table. If the fault persists, contact seca Service or an autho-
rized service partner.
Fault Cause Remedy
Scale shows no weight although
there was a load on it
The device is not ready for use Remove transport lock
Scale does not show a plausible
weight value
Before weighing pointer does not read
0
Adjust display
Force shunt: scale is e.g. touching a
wall, or patient is supporting his/her
weight against the wall
Move scale away from wall
Use transport lock as wall spacer
Ask patient to stand on the scale un-
aided
9 SERVICING
9.1 Non-verified scales
The product needs to be set up carefully and serviced regularly. Depending on
how frequently the product is used, seca recommends servicing at intervals of 3
to 5 years.
CAUTION!
Faulty measurements as a result of poor servicing
Have servicing and repairs carried out exclusively by seca Service
or an authorized service partner.
You can find your local service partner at www.seca.com.
9.2 Verified scales
CAUTION!
Faulty measurements as a result of verification being performed poorly
or not at all
Have verification performed only by authorized persons.
Always have verification performed if one or more quality seals are
damaged.
seca recommends having your device serviced prior to verification.
CAUTION!
Faulty measurements as a result of poor servicing
Have servicing and repairs carried out exclusively by seca Service
or an authorized service partner.
You can find your local service partner at www.seca.com.