32 •
4.3 Fitting the wall spacer
If you wish to position the scale close to a wall, we
recommend using the transport lock (see section
“Removing the transport lock”) as a wall spacer. This
stops the weighing platform from touching the wall and
the measured result from being incorrect.
1. Position the transport lock as shown in the adjacent
drawing.
2. Attach the transport lock to the lower section of the
housing.
4.4 Aligning the scale
ATTENTION!
Incorrect measurement due to force shunts
Soft floor coverings such as carpets may cause
weights to be measured incorrectly.
► Set the scale up so that only its feet are in
contact with the floor.
► Place the scale on a firm, flat surface.
5. WEIGHING
CAUTION!
Bodily injury
Perform a function check before every use as
described in the section “Function check” on
page 36.