Technical data • 39
10.2 Technical data, weighing
Air pressure
• Operation
• Storage
•Transport
700 - 1060 hPa
700 - 1060 hPa
700 - 1060 hPa
Humidity
• Operation
• Storage
•Transport
30 % - 80 % no condensation
0 % - 95 % no condensation
0 % - 95 % no condensation
Medical device in accordance with
Directive 93/42/EEC
Class I with measuring function
General technical data
Technical data, weighing seca 761
Calibration in accordance with
Directive 2014/31/EU
Class IIII
Maximum load 150 kg
Minimum load 10 kg
Graduations 1 kg
Zero adjustment range 20 kg
Accuracy on initial calibration:
• 10 kg to 50 kg
• 50 kg to 150 kg
±0.5 kg
±1 kg
Technical data, weighing seca 750/760/762
Maximum load 150 kg / 320 lbs / 23 sts
Minimum load 10 kg / 20 lbs / 1 sts
Graduations
seca 750/760 1 kg / 1 lbs / 1 lbs
Graduations
seca 762 0,5 kg / 1 lbs / 1 lbs
Zero adjustment range 20 kg / 45 lbs / 3 sts
Accuracy:
• 10 kg to 80 kg
• 80 kg to 150 kg
• 20 lbs to 180 lbs
• 180 lbs to 320 lbs
• 1 sts to 12 sts
• 12 sts to 23 sts
±1 kg
±1.25 %
±2.2 lbs
±1.25 %
±0.2 sts
±1.25 %