What do I do if …? • 37
CAUTION!
Bodily injury
If you find faults or deviations during the
function check which you are unable to remedy
with the aid of the section “What do I do if …?”,
you may not use the device.
► Have the device repaired by seca service or
an authorized service partner.
► You should also follow the instructions in the
section “Maintenance/verification” on
page 37.
8. WHAT DO I DO IF …?
9. MAINTENANCE/VERIFICATION
9.1 Information on servicing (seca 750/760/762)
The product must be set up carefully and serviced
regularly. Depending on how frequently the scale is
used, we recommended servicing at intervals of 3 to 5
years.
ATTENTION!
Incorrect measurements as a result of poor
servicing
►
Please only have servicing and repairs per-
formed by an authorized service partner.
► You can find service partners in your area at
www.seca.com or by sending an e-mail to
service@seca.com.
Malfunction Cause/Remedy
... the scale does not
show any weight?
The device is not ready for use
- Remove transport lock
... the scale does not
show any plausible
weight?
• Before weighing pointer does not read “0”
- Adjust display
• Force shunt: scale is e.g. touching a wall or
patient is supporting his/her weight against the
wall.
- Move scale away from wall
- Use transport lock as wall spacer
- Ask patient to stand on the scale unaided