64 •
ATTENTION!
Malfunction due to assembly error
If the armrest is positioned incorrectly before
assembly, it cannot be folded down after
assembly.
Position the armrest as depicted in the figure.
1. Insert the bolt for the armrest through the hinge
sleeve of the seat frame.
2. Screw the armrest to the seat frame (1x box nut
M10 with washer 10.5 mm).
3. Repeat steps 1. and 2. for the second armrest.
5.4 Establishing the power supply
Power is supplied to the scale with a rechargeable bat-
tery block or a power pack (both included in the scope
of delivery).
Insert the rechargeable
battery block
1. Press on latch of battery compartment.
2. Move down lid of battery compartment.
3. Remove the connector cable from the battery com-
partment.
4. Connect the connector cable to the rechargeable
battery block.
5. Insert the rechargeable battery block into the bat-
tery compartment.
6. Close the battery compartment.