8. Storage, transport and
disposal
8.1 Storage, e.g. Winterization
To place the UrSpring in a (temporary) storage:
Disconnect the electricity and water supply.
Remove water from the UrSpring. Water can become contaminated if it is
not refreshed for a long time.
Clean the external and internal surfaces of the UrSpring and accessories.
Make sure the environment temperature does not drop below 0°C (32°F)
or rise above 40°C (104°F).
Make sure the relative humidity of the environment is below 80%.
Make sure the UrSpring stays in an upright position.
After removing the UrSpring from storage, check the UrSpring for
damage.
Storing filter modules:
Store filter modules in a cool and dry place.
Do not expose filter modules to direct sunlight.
Do not store filter modules for longer than 2 years after the production
date.
Do not remove the packaging before storage.
Before storing an already used filter module, rinse it with a solution of pure water with
1% sodium bisulfite.
If a used filter module is stored longer than 6 months, make sure to thoroughly clean and
rinse the filter with clean drinking water before use.
Storing used filters:
Before storing, rinse the filter module with a solution of pure water with 1%
sodium bisulfite.
Seal filter modules using the original Seccua protective end caps.
Store filter modules vertically with the opening facing upwards.
If a filter module is stored longer than 6 months, make sure to thoroughly clean and rinse
the filter with clean drinking water before use.
Do not freeze!
Page 49/50