The Generator consists of the following essential parts:
• Control Console, operator controls and displays for radiographic
operations. The Console is designed for ease of operation. It is the
interface with the Power Cabinet and other related X-ray systems.
• Power Cabinet, that comprises:
G Power Module, which contains the power and control
components.
G High Voltage Tr ansformer.
1.1 GENERAL FEATURES
The main features of this high frequency Generator are:
• Constant potential high frequency, operating on three phase and single
phase lines.
• Three point control by selecting kVp, mA and Exposure time, or two point
control by selecting kVp and mAs, or one point control by selecting kVp
with AEC operations.
• Two Buckys can be directly connected to the Generator (standard).
• Self-diagnosis indicators identify malfunctions in the system.
• Tube protection circuitry prolongs Tube life and increases system
performance.
• Equipped with closed loop control of X-ray Tube current, kVp and
filaments, which minimize potential errors and the need for
readjustments.
• Automatic line voltage compensation due to closed loop operation of
X-ray Tube current and kVp.
• Independent Heat Unit storage for each X-ray Tube, even after turning
On / Off the equipment.