HF Series Generators -- RAD Console
Operation
OM- 0279R1
18
3.1.4 RADIOGRAPHIC PARAMETERS
kV DISPLAY can show:
The radiographic kV value selected for the technique.
The actual X-ray Tube heat unit value after pressing the On push-button
(Refer to Section 3.6).
The error messages during a system fault, preceded by the letter “E”
(a.e., E02) (Refer to Section 3.9).
mAs DISPLAY can show:
The radiographic mAs value selected for the technique.
When an exposure is made with AEC, it shows the actual mAs at the end
of the exposure whenever the Prep push-button has not been released.
If an exposure is aborted by releasing the Exp push-button during the
exposure, it shows the actual mAs value until the AEC Rese t
push-button is pressed to reset the error condition.
mA DISPLAY: Shows the radiographic mA value selected for the technique.
Time DISPLAY can show:
The Time value (in s econds) selected for the radiographic technique.
When an exposure is made with AEC, it shows the back-up time during
the exposure and the actual Time at the end of the exposure whenever
the Prep push-button has not been released.
If an exposure is aborted by releasing the Exp push-button during the
exposure, it shows the actual Time until the AEC Reset push-button is
pressed to reset the error condition.
RAD Displays canalsoshow:
The actual Time, the calculated mAs, and the selected kVp and mA
radiographic parameters of the last exposure, with or without AEC, after
pressing the AEC Reset push-button (values flashing).
The exposure counters (Refer to Section 3.7).