Table 3-2
Error Codes
ERROR DESCRIPTION WHAT TO DO
-- -- -- -- -- --
on Display
System failure. This indication may appear together with an
error on the Console, and indicates that the error is not
correctable unless the equipment is turned OFF.
Turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
E01, E02 Communication error.
Turn the Generator OFF, check the proper external cable connections
and then turn the Generator ON. If the equipment remains inoperative,
turn it OFF and call Field Service.
E03 System failure.
E04
The Power Cabinet has activated “Preparation” without a
Console command intervention.
Turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
E05 External exposure activated during power-up.
Release any external exposure device or push-buttons.
Turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
E06
“Exposure” or/and “Preparation” orders are activated during
power-up.
Release all the controls.
Turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
E07, E08 X-ray Tube configuration error.
Press the “AEC Reset” push -button.
If the error code persists, turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
E09
Generator Overload error. The exposure has been
interrupted because during exposure has been produced
arcing or bad function on the HV circuitry (X -ray Tube, HV
Transformer and/or HV Cables) or a failure of IGBT module
(overheated or defective IGBTs) has been detected.
It can be also shown making a high power and long
exposure with the X-ray tube cool (X-ray Tube has not been
warmed-up).
This error does not require to press the “AEC Reset” push-button, its
indication disappears automatically.
If the error code persists, turn the Generator OFF and wait 30 minutes
before turning it ON again.
If the equipment remains inoperative, turn it OFF and call Field Service.
E10, E11 System failure.
Press the “AEC Reset” push -button.
If the error code persists, turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
E12 No mA during exposure or mA value is out of range.
Press the “AEC Reset” push -button.
Re
eat with same techni
ue values, I
the error code
ersists tr
with
E13 No kV during exposure or kV value is out of range.
,
another combinations of kV and mA values.
If the equipment remains inoperative, turn it OFF and call Field Service.
E14, E15 System failure.
Press the “AEC Reset” push -button.
If the error code persists, turn the Generator OFF and ON again.
If the equipment remains inoperative, turn it OFF and call Field Service.
E16 Invalid value of: kV, mA or kW.
Decrease kV, mA or both. Press the “AEC Reset” push-button.
If the error code persists, turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
E17 Communication error or system failure.
Turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
E18
Rotor error. The X-ray tube anode is not rotating while
“Prep” is active, then exposures are inhibited, or the X-ray
tube anode is rotating without console command.
This error does not require to press the “AEC Reset” push-button, its
indication disappears automatically.
If the error code persists, turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
E19, E20 System failure.
Turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
E21, E22 Incorrect selection of the X-ray Tube.
“
E23 System failure.
Press the “
EC Rese
” push-button.
If the error code persists, turn the Generator OFF and ON.
E24 Bucky not ready for an exposure.
If the equipment remains inoperative, turn it OFF and call Field Service.