DIGITAL WEIGHING INDICATOR
SI 4010
12. Input “Test weight” capacity. And press key.
Ex) Load Cell CAPA : 20kg, division 0.001
Use test weight(at least 2kg) which is 10% of max CAPA(20kg)
input test weight 2.000
13. After displaying “UP” Load “Test Weight”
Ex) Load Cell CAPA : 20kg, division 0.001
Use test weight(at least 2kg) which is 10% of max CAPA(20kg)
14. And press key.(Do not unload test weight)
15. Indicator will calculate span value during 5sec, automatically
16. After calculation, span value displays on the display.
Please unload the test weight.
※ Caution
The “Test Weight’s value” must be at least 10% max. capacity of weighing scale.
“at least 10%” means to guarantee precise weighing process you have to make standard with at least 10% of
weight of Max. capacity.
We programmed the calibration will not be done, when you load less than 10% of max. capacity.
17. Press key to save all calibration process.
After then it resets automatically.
(Now, fasten the Calibration Bolt. )