Entering means ESC/UPPER step, Entering means SAVE/NEXT Step.
■ Adjusting “ZERO” Balance (Calibration)
Adjust weight balance between “Real weight” on the load cell(Weight Part) and
“Displayed weight of Indicator”. When you replace LOAD CELL or Indicator, you
have to Calibrate process once again.
(When you start calibration mode, TARE, HOLD & PRINT will be reset.)
Before processing calibration, please warm up the indicator during 15
min to guarantee more preciseness.