SI 480E DIN SIZE WEIGHING CONTROLER
5-3. Simulation Calibration Mode(Calibrate without Test weight)
With this Simulation Calibration Mode you can make simple calibration without any TEST weight
This calibration mode uses Load cells max capacity and Max. Output Rate(mV), so the weight
adjustment degree might be less than Test weight Calibration.
The guaranteed resolution of this Simulation Calibration is 1/3,000.
5-3-1. Simulation Calibration Mode Start
Under SEtUP Mode, press key.
CALIbr displays, press key.
If S-CAL displays, press key to start
Simulation Calibration Mode
5etUp
Cal1br
5-Cal