16
SI 480E DIN SIZE WEIGHING CONTROLER
5-2-5. Input Test Weight value and Calculate SPAN value.
If SPAN is displayed, input Test Weight capacity and press key.
If UP is displayed, please load Test Weight on the scale part and press key.
Calculate Span value during 10 ~20 secs,
automatically.
Over than 1/10,000 resolution setting,
To guarantee the preciseness, Span calculation will be operated twice.
After calculation, span value will be displayed
on the display. Then press key.
This span value is not a weight value.
When END is displayed and calibration is
completed.
5paN
1000.0
Up
Cal-10
0.33392
end