SI 480E DIN SIZE WEIGHING CONTROLER
5-2-4. Measure the DEAD Weight of Weighing Scale.
When DEAD displays, press key, then indicator will calculate Dead weight of scale part
automatically.
Indicator will search DEAE weight during 10~20
secs automatically to find the best condition.
Over than 1/10,000 resolution setting,
To guarantee the preciseness, DEAD weight calculation (CAL00~CAL09) will be operated twice.
In this step, if there is some force or Vibration on scale part, these unstable
conditions will be continued, ErrorA will be displayed, and DEAD value will not be
calculated.
Under this condition, please remove the cause of force or vibration and process it again.
dead
Cal-00