SI 480E DIN SIZE WEIGHING CONTROLER
5-2-3. Decimal Point and Digit / Division Value
After DIVI is displayed select Decimal point with key.
Whenever pressing key , decimal point will be changed.
Please stop at the optimal position. And select Division optimal division with key.
Finally press key to save and move to next step.
Max. Decimal point will be 0.001, and digit can be selectable among
1, 2, 5, 10, 20, 50.
Digit and Decimal point must be fulfill the below condition.
- (Max. capacity value / division value) cannot be over than 20,000.
If this condition is not fulfilled, err-1 will be displayed and move back to
Capacity setting mode.
diui
0.1