43
SI 480E DIN SIZE WEIGHING CONTROLER
7-2. Calibration Process
Display
Cause
Treatment
Eerr01
When Max.capacity/digit value is over
20,000
Re-input the Max. Capacity, less than
20.00
(Max. Capacity / Digit)
Eerr04
Standard weight value is over than Max.
Capacity
Re-input Standard weight value with
Number keys, under Max. Capacity
Eerr05
Standard weight value is less than 10% of
Max. Capacity
Re-input Standard weight value with
Number keys, more than 10% of Max.
Capacity
Eerr06
1. Amp. Gain is too big
2. Sig+ and Sig- wire connection error
3. Test weight is not loaded
Check standard weight’s weight with set
value.
If there is difference between set value
and real weight, please re-input the value
(set value is too small)
Eerr07
1. Amp. Gain is too small
2. Sig+ and Sig- wire connection error
3. Test weight is not loaded
Check standard weight’s weight with set
value.
If there is difference between set value
and real weight, please re-input the value
(set value is too big)
Eerr08
Under “F-function” model, set value is
“N.A”
Check the correct value and re-input
err-a
When there is continuous vibration on the
weighing part,, indicator cannot process
calibration any more.
- Find vibration cause and remove
- Load cell check
- Load cell cable and connecting
condition check