SI 510 WEIGHING INDICATOR
5-3-2. Hidden Function Setting Mode
This setting become enter the hidden Function..
How to enter Hidden Function Setting Mode : Press "F2" key 4 times and input your
password. (1111)
* Key function: F1 – Move hidden function number / Enter – Save
Check your device's serial number.
Check the currently applied program version.
Check the currently applied hardware version.
DATE(Y,M,D) Check / Modification
Check the date or adjust when it is wrong.
TIME(H,M,S) Check / Modification (24Hours)
Check the time or adjust when it is wrong.
Password Setting (4digit)
Password is required when you enter to hidden function.
Enter the password twice.
Password combination within 0~9.
Maximum Capacity Weight Check
Check the max capacity which is set under calibration mode.
Span Value Check / Modification
Analog Output Setting Set up
Min Analog Output Setting
Minimum Analog Output (Analog out 4~20mA / 0~10V).
The 4-20mA’s begin number is “0”, so after enter the Function mode and write the “4” to
show “4mA” key press (–) Setting. Input range : -20 ~ +20 , basic value : 0
Max Analog Output Setting
Maximum Analog Output (Analog out 4~20mA / 0~10V).
The 4-20mA’s begin number is “0”, so after enter the Function mode and write the “-4” to
show “20mA”
key press (–) Setting. Input range : -20 ~ +20 , basic value : 0
Change the function list reset.
Change the all function to the factory reset.
Ip Address Check / Modification