M054-E093J
Instruction Manual
for
Electronic Balance
AW series AW320, AW220, AW120
AX series AX200, AX120
AY series AY220, AY120, AY62

Table of Contents

Questions and Answers

  • R
    Robert WalkerAug 9, 2025
    What does 'CAL E2' mean on my Shimadzu Scales?
    • Y
      yhallAug 9, 2025
      If your Shimadzu Scales displays 'CAL E2', it means something was on the pan during span calibration. Remove the item and wait for the span calibration to restart.
  • K
    Kathy WilkinsAug 15, 2025
    How to troubleshoot a Shimadzu Scales with a blank display?
    • C
      Cindy WatersAug 15, 2025
      If nothing appears on the display of your Shimadzu Scales, first ensure that the AC adapter is properly connected. If it is, check that the power supply is turned on.
  • A
    Amanda MillerAug 20, 2025
    Why does the display on my Shimadzu Scales fluctuate?
    • J
      Jordan TerryAug 21, 2025
      If your Shimadzu Scales display is fluctuating, here are a few things to consider: * The balance might be experiencing vibration or drafts. Try relocating it. * If weighing a volatile sample, use a lid. * If the sample is charged, place it in a metallic container. * Ensure the sample's temperature matches the weighing chamber's. * Separate the balance from any electrical noise or electromagnetic waves. * If air currents are present, keep the weighing chamber door slightly open when not in use.
  • E
    Elizabeth ChambersAug 27, 2025
    What to do if my Shimadzu Scales displays 'Err 20'?
    • C
      Cassidy AlvaradoAug 27, 2025
      If the Shimadzu Scales displays 'Err 20', it indicates that an incorrect value has been entered. Review the numeric value you entered and correct it.
  • V
    Virginia BrownSep 2, 2025
    What does 'Err24' mean on my Shimadzu Scales and how do I fix it?
    • L
      Lance SmithSep 3, 2025
      If your Shimadzu Scales displays 'Err24', this indicates low battery pack voltage. You should charge the battery pack.

Summary

Safety Precautions

2. INSTALLATION

3. WARM-UP

5. MEASUREMENT PROCEDURE

6. MENU SELECTION

6.1 Operation Menu

7. SPAN CALIBRATION (Sensitivity Adjustment)

7.2 Span Calibration Using an External Weight

10. REGISTRATION, RELEASE, AND SELECTION OF MEASURE-MENT UNITS

10.1 Percentage

13. ADD LOAD WEIGHING

14. STANDARD SPAN CALIBRATION (Sensitivity Adjustment)

17. CALIBRATION OF THE BUILT-IN WEIGHT (AW;AX series only)

20. AUTOMATIC SPAN CALIBRATION (PSC)(AW series only)

30. TROUBLESHOOTING

30.1 Error Message

31. SPECIFICATIONS

34. Windows DIRECT FUNCTION

36. COMMAND CODE

Shimadzu AY120 Specifications

General IconGeneral
Capacity120 g
Readability0.1 mg
Pan Size80 mm diameter
DisplayLCD
Repeatability0.1 mg
Linearity±0.2 mg
Operating Temperature5°C to 40°C
InterfaceRS-232C
CalibrationExternal
Response TimeApproximately 3 seconds
Power SupplyAC adapter