M054-E093J
Instruction Manual
for
Electronic Balance
AW series AW320, AW220, AW120
AX series AX200, AX120
AY series AY220, AY120, AY62

Table of Contents

Questions and Answers

  • N
    Nicole EdwardsAug 9, 2025
    What does 'CAL E2' mean on my Shimadzu Scales?
    • M
      Michael TannerAug 9, 2025
      If your Shimadzu Scales displays 'CAL E2', it means something was on the pan during span calibration. Remove the item and wait for the span calibration to restart.
  • M
    Maureen BoyerAug 15, 2025
    How to troubleshoot a Shimadzu Scales with a blank display?
    • G
      Glenn CombsAug 15, 2025
      If nothing appears on the display of your Shimadzu Scales, first ensure that the AC adapter is properly connected. If it is, check that the power supply is turned on.
  • D
    daviskennethAug 21, 2025
    Why does the display on my Shimadzu Scales fluctuate?
    • K
      Kevin SmithAug 22, 2025
      If your Shimadzu Scales display is fluctuating, here are a few things to consider: * The balance might be experiencing vibration or drafts. Try relocating it. * If weighing a volatile sample, use a lid. * If the sample is charged, place it in a metallic container. * Ensure the sample's temperature matches the weighing chamber's. * Separate the balance from any electrical noise or electromagnetic waves. * If air currents are present, keep the weighing chamber door slightly open when not in use.
  • R
    reginabrowningAug 27, 2025
    What to do if my Shimadzu Scales displays 'Err 20'?
    • A
      Andre PriceAug 27, 2025
      If the Shimadzu Scales displays 'Err 20', it indicates that an incorrect value has been entered. Review the numeric value you entered and correct it.
  • A
    Ashley ChavezSep 1, 2025
    What does 'Err24' mean on my Shimadzu Scales and how do I fix it?
    • A
      Anita OrtegaSep 1, 2025
      If your Shimadzu Scales displays 'Err24', this indicates low battery pack voltage. You should charge the battery pack.

Summary

Safety Precautions

1. BALANCE COMPONENTS

Key Functions

2. INSTALLATION

Installation Procedure

Step-by-step guide for setting up the balance, including power and leveling.

3. WARM-UP & PRECAUTIONS

5. MEASUREMENT PROCEDURE

Measurement

Explains the process of taking measurements, including taring and reading results.

6. MENU SELECTION

6.2-6.6 Menu Navigation

Overview of the balance's menu structure and navigation for settings and functions.

14. SPAN CALIBRATION (Sensitivity Adjustment)

17. BUILT-IN WEIGHT CALIBRATION (AW;AX series only)

28. PERFORMANCE CHECK

Repeatability & Corner Load Error

30. TROUBLESHOOTING

30.1 Error Message

Lists common error messages and their corrective actions.

31. SPECIFICATIONS

32. PARTS LIST

34. Windows DIRECT FUNCTION

35. DATA FORMATS

36. COMMAND CODE

Shimadzu AY220 Specifications

General IconGeneral
BrandShimadzu
ModelAY220
CategoryScales
LanguageEnglish