Gas Chromatograph
GC-8A Series
C184-E016B
Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-8AIT and is the answer not in the manual?

Shimadzu GC-8AIT Specifications

General IconGeneral
TypeGas Chromatograph
WeightApprox. 50 kg
Column TypePacked or Capillary
Detector TypeFID, TCD, ECD, FPD
DetectorFID, TCD, ECD, FPD
Carrier GasNitrogen, Helium, Argon, Hydrogen
Temperature RangeAmbient +5°C to 399°C
Oven TemperatureAmbient +5°C to 399°C
Column Oven Temperature RangeAmbient +5°C to 399°C

Summary

Column Oven

Rapid heating and cooling characteristics

Column oven heats from 100°C to 350°C in 12 mins and cools in under 1 min.

Two-stage overheat prevention

Heater turns off if temperature exceeds preset by 30°C or reaches 420°C.

Temperature control stability and response

Excellent temperature control is essential, showing rapid response to changes.

Detectors

Thermal Conductivity Detector (TCD)

Features highly stable, sensitive filaments and a constant current system for TCD.

Flame Ionization Detector (FID)

Utilizes a cylindrical collector and on-detector system for low contamination in FID.

Electron Capture Detector (ECD)

Linear ECD with high sensitivity, wide dynamic range, and switchable detector current.

Flame Photometric Detector (FPD)

Detects sulfur and phosphorus selectively; features FID monitor electrode and forced cooling.

Accessories