J
Jeanette FosterJul 30, 2025
What to do if ignition failed on Shimadzu GC-2010 Plus Laboratory Equipment?
- NnrussellJul 30, 2025
If the ignition fails and the message “Ignition failed.” appears, consider these potential causes and solutions: * The column might not be connected: Connect the column for FID to be ignited and ensure that carrier gas is flowing. * Gas leaks at the detector side column connection: Tighten the nut or replace the ferrule. * Hydrogen or air is not supplied: Supply hydrogen or air. * Incorrect hydrogen or air flow rate: Set the hydrogen or air flow rate to a proper value. * Unused FID/FPD is set to “On”: Set unused detectors to “Off.” * FID jet is clogged: Clean or replace the FID jet. * Igniter filament is broken: Replace the igniter.