Shimadzu GC-2010 Plus - User Manual

Shimadzu GC-2010 Plus
352 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
221-40803
GC-2010 Plus
Gas Chromatograph
INSTRUCTION MANUAL
Read the instruction manual thoroughly before you use the product.
Keep this instruction manual for future reference.

Table of Contents

Questions and Answers

  • J
    Jeanette FosterJul 30, 2025
    What to do if ignition failed on Shimadzu GC-2010 Plus Laboratory Equipment?
    • N
      nrussellJul 30, 2025
      If the ignition fails and the message “Ignition failed.” appears, consider these potential causes and solutions: * The column might not be connected: Connect the column for FID to be ignited and ensure that carrier gas is flowing. * Gas leaks at the detector side column connection: Tighten the nut or replace the ferrule. * Hydrogen or air is not supplied: Supply hydrogen or air. * Incorrect hydrogen or air flow rate: Set the hydrogen or air flow rate to a proper value. * Unused FID/FPD is set to “On”: Set unused detectors to “Off.” * FID jet is clogged: Clean or replace the FID jet. * Igniter filament is broken: Replace the igniter.
  • N
    Nicolas KimAug 9, 2025
    Why does the message “Ignition failed” appear on my Shimadzu Laboratory Equipment?
    • P
      Penny GarnerAug 9, 2025
      The “Ignition failed” message on your Shimadzu Laboratory Equipment may appear due to several reasons: * The column is not connected. * Gas leaks at the detector side column connection. * Hydrogen is not supplied, or the hydrogen flow rate is incorrect. * Air is not supplied, or the air flow rate is incorrect. * Unused FID/FPD is set to “On”. * FID jet is clogged. * Igniter filament is broken. Check these potential causes and apply the suggested solutions.
  • C
    charlesfinleyAug 18, 2025
    How to fix 'Ignition failed' error on Shimadzu Laboratory Equipment?
    • L
      Lori GonzalezAug 19, 2025
      To resolve the 'Ignition failed' error on Shimadzu Laboratory Equipment, you can: 1. Connect the column for FID and ensure carrier gas flow. 2. Tighten the nut or replace the ferrule if there are gas leaks. 3. Supply hydrogen and air if they are not being supplied. 4. Adjust the hydrogen and air flow rates to the correct values. 5. Turn off any unused detectors. 6. Clean or replace the FID jet if it's clogged. 7. Replace the igniter if the filament is broken.

Summary

WARNING LABELS

WARNING Label Convention

Explains the meaning and implications of the "WARNING" label.

CAUTION Label Convention

Explains the meaning and implications of the "CAUTION" label.

NOTE Label Convention

Explains the meaning and purpose of the "NOTE" label.

Safety Precautions

General Safety Guidelines

Lists essential safety rules for operating the GC-2010 Plus.

System WARNING Labels

Hot Surfaces and Fire Hazard

Warns about hot components (ports, cover) and fire risks.

Rear Panel Burn Hazard

Warns about the hot rear panel potentially causing burns.

Hydrogen Use and Explosion Hazard

Advises on safe hydrogen handling to prevent explosions.

Hot Air Exhaust Warning

Advises keeping heat-sensitive materials away from exhaust openings.

High Voltage Hazard

Warns of high voltage hazards and recommends qualified personnel for service.

Shimadzu GC-2010 Plus Specifications

General IconGeneral
TypeGas Chromatograph
ModelGC-2010 Plus
ManufacturerShimadzu
Maximum Temperature Ramp Rate120 °C/min
Detector OptionsFID, TCD, ECD, FPD, NPD, MS
Electronic Flow Control (EFC)Yes
Data Sampling RateUp to 250 Hz
Column Oven Temperature RangeAmbient +5°C to 450°C
Maximum Oven Heating Rate120 °C/min
Injection PortSplit/Splitless, PTV
DisplayLCD

Related product manuals