1 Overview
1.2
1.
2
GC-2010 Plus
1.2Specifications
1.2.1 Column oven
Dimensions (mm) : 280 (W) × 280 (H) × 175 (D) (excluding Projections)
Volume (L) : 13.7
Range of temperature : Room temperature +4 °C to 450 °C
-50 °C to 450 °C (When liquid carbon dioxide gas is used.)
Accuracy of temperature : ±1 % (K) (Calibrated at 0.01 °C)
Deviation of temperature : Within 2 °C (on a 200 mm diameter column holder)
Stability of temperature : Within ±0.05 °C
Temperature coefficient : 0.01 °C/ °C
Range of linear temperature increase:
(in power voltage 100 VAC)
40 °C/min up to 200 °C
25 °C/min up to 250 °C
15 °C/min up to 380 °C
7 °C/min up to 450 °C
(in power voltage 230 VAC)
70 °C/min up to 200 °C
50 °C/min up to 350 °C
35 °C/min up to 450 °C
Cooling speed : Approximately 3.4 minutes cooling from 450 °C to 50 °C.
(Ambient temperature : 22 °C)
Overheat protection : Programmable up to 470 °C (A fixed circuit provides protection
at 500 °C)
1.2.2 Temperature program
Program ramps : 20 ramps in total (Heating and cooling available)
Setting : 0.1 °C increments
Program setting : -250 to 250 °C/min, 0.01 °C/min increments
Total time of total program : Up to 9999.99 minutes
1.2.3 Injection port
Overheat protection : Programmable up to 470 °C
SPL-2010 Plus
Injection unit : Split/Splitless injection
Range of temperature : Up to 450 °C
Temperature setting : 0.1 °C increments
WBI-2010 Plus
Injection unit : Direct injection
Range of temperature : Up to 450 °C
Temperature setting : 0.1 °C increments