SHIMADZU
Analytical Balance
Instruction Manual
321-78382
Oct. 2021
Read the instruction manual thoroughly before you use the product.
Keep this instruction manual for future reference.
1
3
5
2
4
6
7
8
12
13
9
10
11
Foreword
Table of Contents
Setup
Component Names
and Functions
Display and Icons
Basic Operations
Calibration
Settings
Applications
Ionizer
Connecting and Communicating
with Peripheral Devices
Errors and Recovery
Maintenance
Specifications
Technical
Documentation
AP W Series
AP225W
AP135W
AP225WD
AP125WD
AP324W
AP224W
AP124W
AP X Series
AP324X
AP224X
AP124X
AP Y Series
AP324Y
AP224Y
AP124Y
AP W-AD Series
AP225W-AD
AP135W-AD
AP225WD-AD
AP125WD-AD
AP324W-AD
AP224W-AD
POWER
ION
ALARM

Table of Contents

Questions and Answers

  • S
    Susan RandolphAug 10, 2025
    What to do if the serial communication function doesn't work with a USB serial converter on Shimadzu Laboratory Equipment?
    • L
      Lisa BlevinsAug 10, 2025
      First, verify that the correct COM number has been set in the Windows settings. If the issue persists, the driver provided with the USB serial converter may not be set up correctly. Please refer to the Shimadzu website for details on checking/changing the COM number and setting up the driver.
  • M
    Mathew GreeneAug 15, 2025
    Why doesn't the serial communication function work after rebooting the PC with Shimadzu AP324X AP224X Laboratory Equipment?
    • C
      cathy55Aug 16, 2025
      The serial communication function cannot be used if Balance Keys has not been started up. For details on Balance Keys, refer to the Shimadzu website.
  • K
    Kevin WeberAug 21, 2025
    What to do if the cells are not transferred when entering data on Excel with Shimadzu AP324X AP224X Laboratory Equipment?
    • D
      danielwilliamsAug 21, 2025
      This issue may occur if Excel is not set to move cells after pressing the Enter key. To resolve this, in Excel, go to File > Options > Advanced Settings, and enable 'Move cell after pressing the [Enter] key.' Also, ensure that the Japanese conversion function in Windows is set to OFF, as this can interfere with data transfer.
  • J
    Joseph HicksAug 27, 2025
    Why do operations sometimes seem strange with Shimadzu Laboratory Equipment?
    • C
      cooperjenniferAug 27, 2025
      Strange operations can occur if the communication speed is set too slow in Balance Keys, depending on your PC's performance. Adjust the communication speed to align with your PC's capabilities. Also, if the interval for transmitting data from the balance is too short, it can cause malfunctions. Ensure data is transmitted after it has been displayed on the screen and avoid using the continuous output function. Additionally, ensure nothing is touching the PC keyboard or mouse when data is being transmitted from the balance.
  • L
    Laura FarleySep 3, 2025
    What to do if garbled text is loaded onto the PC from Shimadzu AP324X AP224X?
    • C
      Caitlin MitchellSep 3, 2025
      If garbled text appears on your PC, the serial communication function settings on either the balance or the PC are likely incorrect. Reset these settings to resolve the issue.

Summary

Safety Precautions

Precautions on Use

Covers usage, installation, work/operations, and general safety advice.

Risks in Repairs;Dismantling;Modifications

Details risks associated with disassembling, modifying, or repairing the product.

Inspection and Maintenance Precautions

Guidelines for safe inspections, maintenance, and handling during emergencies/outages.

1 Setup

Installation Environment and Parts Attachment

Details selecting an installation site and attaching initial balance components.

Balance Leveling and Power On Procedure

Steps for adjusting the balance's horizontal level and powering it on.

2 Component Names and Functions

Main Unit Front, Top, and Side Panels

Identifies and explains components on the front, top, and side of the balance unit.

4 Basic Operations

Standard Measurement and Ending Operations

Guides users through basic weighing procedures and how to end measurements.

5 Calibration

CAL Key Settings and Calibration Methods

Explains CAL key setup and the available methods for balance calibration.

Internal and External Weight Adjustments;Tests

Details procedures for calibrating and testing with internal and external weights.

Perfect Self Calibration and Timer CAL

Covers automatic sensitivity adjustment (PSC) and time-scheduled calibration (Timer CAL).

Periodic Inspections and Error Checks

Guides on performing periodic inspections and basic error checks.

6 Settings

7 Applications

10 Errors and Recovery

General Troubleshooting and Error Messages

Provides solutions for common problems and explains various error messages.

11 Maintenance

General Balance and Ionizer Maintenance

Instructions for cleaning and maintaining the balance unit and its ionizer.

Inspections and Weight Management

Guidelines for daily/periodic inspections and selecting/managing calibration weights.

Shimadzu AP324X AP224X Specifications

General IconGeneral
Capacity (AP324X)320 g
Capacity (AP224X)220 g
Readability0.1 mg
Linearity (AP324X)±0.2 mg
Linearity (AP224X)±0.2 mg
InterfaceRS-232C
ModelAP324X AP224X
Repeatability (AP324X)0.1 mg
Repeatability (AP224X)0.1 mg
Stabilization Time2 seconds
DisplayLCD
Power SupplyAC adapter
Operating Temperature5 to 40 °C
Pan SizeØ 80 mm