1 Overview
1.2 Specifications
4
GC-2010 Plus
1.2.6 Carrier gas flow control unit
○ Split/splitless mode
Range : 0 to 970 kPa (The maximum pressure limit is the primary pres-
sure minus 10 kPa.)
0.1 kPa increments
Program ramps : 7 ramps possible
Program rate : -400 to 400 kPa/min, 0.01 kPa/min increments
Split rate setting : 0 to 9999.9, 0.1 increments
○ Direct injection mode
Pressure mode
Range : 0 to 970 kPa (The maximum pressure limit is the primary pres-
sure minus 10 kPa.)
0.1 kPa increments
Program ramps : 7 ramps possible
Program rate : -400 to 400 kPa/min
Flow rate mode
Range : 0 to 1200 ml/min (When primary pressure is 980 kPa)
Program ramps : 7 ramps possible
Program rate : -400 to 400 ml/min/min
0.1 ml/min/min increments
1.2.7 Detector gas flow controller
Range : 0 ~ 1200 ml/min (Air), 0.1 ml/min increments
0 ~ 100 ml/min (H
2)
0 ~ 100 ml/min (Makeup He)
Program ramps : 7 ramps possible
Program rate : -400 to 400 ml/min/min, 0.01 ml/min/min increments
1.2.8 Display
Back-light LCD 240 × 320 dot, 16 lines
The display can be switched between Japanese and English.
1.2.9 Dimensions, weight and power supply
Dimensions (mm) : 515 (W) × 440 (H) × 530 (D) mm
Weight : 30 kg (standard model with FID detector)
Power supply : 100 VAC (standard model with FID detector), 1800 VA, 50/60 Hz
230 VAC (standard model with FID detector), 2600 VA, 50/60 Hz